Battery Lifetime
The battery lifetime is limited and varies according to circumstances. In
normal conditions battery pack lasts three years. Esaote recommends to
replace the battery pack every three years.
NOTE
The battery pack has to be replaced by Esaote personnel. This person will
be responsible for ensuring that the system is working properly.
Error Messages
Whenever an internal fault occurs, the system automatically freezes and an
error message is displayed on the screen. Switch the system off and then turn
it on again to see whether the error message persists.
AO
Save anyway the log file (refer to the “Archive” section of the Advanced
Operation manual for further information) and contact the Esaote Service
department.
Errors in Battery Management
The battery icon is shown crossed out whenever an error in the battery
management occurs.
The number in the warning message indicates the type of error.
Error #1
This error indicates a fault on the Power Supply: in this case, information on
batteries may not be correct. The system displays the following message:
If this situation occurs, shut down
MyLab
, by keeping
O N /O FF
pressed, and
contact Esaote Service.
Error #2
This error indicates a failed access to the battery pack: in this case,
information on batteries may not be correct. The system displays the
following message:
If this situation occurs, shut down
MyLab
, by keeping
O N /O FF
pressed, and
contact Esaote Service.
Error #1: wrong communication with
power supply. The automatic shut down
is disabled.
Error #2: wrong communication with
battery logics. The automatic shut down
is disabled.
Summary of Contents for MyLabOmega
Page 1: ...Rev 04 July 2019 MyLabOmega GETTING STARTED 350035100...
Page 2: ...MyLab G E T T I N G S T A R T E D ii...
Page 5: ...MyLab G E T T I N G S T A R T E D v GETTING STARTED EC Declaration of Conformity...
Page 6: ...MyLab G E T T I N G S T A R T E D vi RED Declaration of Conformity...
Page 7: ...MyLab G E T T I N G S T A R T E D vii GETTING STARTED...
Page 8: ...MyLab G E T T I N G S T A R T E D viii...
Page 14: ...I N D E X MyLab G E T T I N G S T A R T E D xiv...
Page 26: ...I N T R O D U C T I O N MyLab G E T T I N G S T A R T E D 1 12...
Page 52: ...S Y S T E M O V E R V I E W MyLab G E T T I N G S T A R T E D 3 18...
Page 62: ...P R E P A R I N G T H E S Y S T E M MyLab G E T T I N G S T A R T E D 4 10...