23
/ 42
5
•
Remedying malfunctions
801
16-
221
10.
15
20
The CUT footswitch pedal was
pressed when switching on the unit.
Do not press the CUT footswitch
pedal when switching on the
unit.
Check the correct function of
the footswitch.
If the error still occurs, notify
Erbe Service.
21
The COAG footswitch pedal was
pressed when switching on the unit.
Do not press the COAG
footswitch pedal when switch-
ing on the unit.
Check the correct function of
the footswitch.
If the error still occurs, notify
Erbe Service.
22
The CUT fingerswitch was pressed
when switching on the unit.
Do not press the CUT finger-
switch when switching on the
unit.
Check the correct function of
the fingerswitch.
If the error still occurs, notify
Erbe Service.
23
The COAG fingerswitch was pressed
when switching on the unit.
Do not press the COAG finger-
switch when switching on the
unit.
Check the correct function of
the fingerswitch.
If the error still occurs, notify
Erbe Service.
24
The selection button (CUT) was
pressed when switching on the unit.
Check whether the correspond-
ing button was pressed when
switching on the unit.
If not, notify Erbe Service.
25
Only for V 1.0.0: The selection button
(COAG) was pressed when switching
on the unit.
Check whether the correspond-
ing button was pressed when
switching on the unit.
If not, notify Erbe Service.
26
The selection button for activation
type was pressed when switching on
the unit.
Check whether the correspond-
ing button was pressed when
switching on the unit.
If not, notify Erbe Service.
Error
code
Description
Additional informa-
tion
Action
Summary of Contents for VIO 100 C
Page 1: ...10 15 80116 221 SERVICE MANUAL VIO 100 C VIO 50 C V 1 0 x V 1 0 x ELECTROSURGERY...
Page 2: ......
Page 3: ...SERVICE MANUAL VIO 100 C VIO 50 C...
Page 6: ...Table of Contents 6 42 80116 221 10 15...
Page 20: ...4 Service mode 20 42 80116 221 10 15...
Page 26: ...5 Remedying malfunctions 26 42 80116 221 10 15...