4
•
Service mode
14
/ 42
801
16-
221
10.
15
2
Volume for
COAG activa-
tion
Volume setting of
the COAG activation
tones in 10 levels.
Activate setup:
Press the yellow selection button.
The current setting appears in the
blue display.
Change settings:
Press the blue Plus/Minus buttons.
The setting change appears in the
blue display.
Save setting change: Press and hold the top left selection
button until a check mark in the
adjacent display indicates the end of
the process.
Exit setup:
Press the yellow selection button
again. The settings display goes out
in the blue display.
3
Volume but-
ton
Volume setting of
the buttons in 10
levels.
Activate setup:
Press the yellow selection button.
The current setting appears in the
blue display.
Change settings:
Press the blue Plus/Minus buttons.
The setting change appears in the
blue display.
Save setting change: Press and hold the top left selection
button until a check mark in the
adjacent display indicates the end of
the process.
Exit setup:
Press the yellow selection button
again. The settings display goes out
in the blue display.
4
Warning tone
volume
Volume setting of
the warning tones in
8 levels.
Activate setup:
Press the yellow selection button.
The current setting appears in the
blue display.
Change settings:
Press the blue Plus/Minus buttons.
The setting change appears in the
blue display.
Save setting change: Press and hold the top left selection
button until a check mark in the
adjacent display indicates the end of
the process.
Exit setup:
Press the yellow selection button
again. The settings display goes out
in the blue display.
No.
Designation
Description
Use
Summary of Contents for VIO 100 C
Page 1: ...10 15 80116 221 SERVICE MANUAL VIO 100 C VIO 50 C V 1 0 x V 1 0 x ELECTROSURGERY...
Page 2: ......
Page 3: ...SERVICE MANUAL VIO 100 C VIO 50 C...
Page 6: ...Table of Contents 6 42 80116 221 10 15...
Page 20: ...4 Service mode 20 42 80116 221 10 15...
Page 26: ...5 Remedying malfunctions 26 42 80116 221 10 15...