Maintenance and servicing
801
16-
221
10.
15
SOFT COAG
FORCED COAG
Test setup
Fig. 6-4
•
The test specimen is connected to the power supply via the power
cord.
•
The levels of power are determined with the HF power meter. The
measuring cables are plugged into the HF power meter direct.
Test procedure for VIO 50 C
1. Set the test specimen to:
SOFT COAG, 50 Watt
2. Set HF power meter to:
RL = 100 Ohm
3. Activate test specimen via COAG button on the electrode handle.
4. Determine and document measured value. The tolerance range is 40
to 60 watts.
1. Set the test specimen to:
FORCED COAG, 50 Watt
2. Set HF power meter to:
RL = 500 ohms
3. Activate test specimen via COAG button on the electrode handle.
4. Determine and document measured value. The tolerance range is 40
to 60 watts.
20196-045
Bipolar cable
or
Bipolar cable, international 2PIN 22
20196-053
20100-034
Adapter cable bipolar
–
HF power meter
20189-107
Two pedal footswitch VIO C
Erbe Art. No.
Description
AE
NE
HF power meter
+
+