Maintenance and servicing
801
16-
221
10.
15
Fingerswitch activation
Testing and measuring equipment
Test setup
•
The test specimen is connected to the power supply via the power
cord.
Test procedure
Electrode handle
Activation CUT / COAG
1. Connect the electrode handle with patient cable AE to the Monopolar
receptacle of the test specimen.
2. Press the CUT button on the electrode handle. The "X" character must
appear in the left segment in the COAG display
2
during activation.
3. Press the COAG button on the electrode handle. The "X" character
must appear in the right segment in the COAG display during activa-
tion.
4. Remove the electrode handle.
Exit test program
1. Switch the test specimen off.
AUTO START mode (for VIO 100 C only)
Testing and measuring equipment
Erbe Art. No.
Description
20192-127
Patient cable AE
or
Patient cable AE, international
20192-110
20190-045
Electrode handle ICC/ACC
2. Legend: "–" not yet activated, "X" successful activation, "0" = activation finished
Erbe Art. No.
Description
20196-045
Bipolar cable
or
Bipolar cable, international 2PIN 22
20196-053
20100-034
Adapter cable bipolar
20100-102
VIO Testbox Auto Start/Auto Stop (bipolar
start/stop)