Test procedure for VIO 100 C
1. Set the test specimen to:
SOFT COAG, 80 Watt
2. Set HF power meter to:
RL = 100 Ohm
3. Activate test specimen via COAG button on the electrode handle.
4. Determine and document the measured value. The tolerance range is
64 to 96 watts.
1. Set the test specimen to:
FORCED COAG, 50 Watt
2. Set HF power meter to:
RL = 500 ohms
3. Activate test specimen via COAG button on the electrode handle.
4. Determine and document measured value. The tolerance range is 40
to 60 watts.
BIPOLAR SOFT COAG
Test setup
Fig. 6-5
•
The test specimen is connected to the power supply via the power
cord.
•
The levels of power are determined with the HF power meter. The
measuring cables are plugged into the HF power meter direct.
•
The dual-pedal footswitch is connected.
Test procedure for VIO 50 C
1. Set the test specimen to:
BIPOLAR (=BIPOLAR SOFT COAG), 50 Watt
2. Set HF power meter to:
RL = 100 Ohm
3. Activate test specimen via COAG pedal on the footswitch.
4. Determine and document measured value. The tolerance range is 40
to 60 watts.
HF power meter