5
•
Remedying malfunctions
24
/ 42
801
16-
221
10.
15
27
The Plus selection button (CUT) was
pressed when switching on the unit.
Check whether the correspond-
ing button was pressed when
switching on the unit.
If not, notify Erbe Service.
28
The Minus selection button (CUT) was
pressed when switching on the unit.
Check whether the correspond-
ing button was pressed when
switching on the unit.
If not, notify Erbe Service.
29
The Plus selection button (COAG) was
pressed when switching on the unit.
Check whether the correspond-
ing button was pressed when
switching on the unit.
If not, notify Erbe Service.
30
The Minus selection button (COAG)
was pressed when switching on the
unit.
Check whether the correspond-
ing button was pressed when
switching on the unit.
If not, notify Erbe Service.
31
The focus button for bipolar recepta-
cle was pressed when switching on
the unit.
Check whether the correspond-
ing button was pressed when
switching on the unit.
If not, notify Erbe Service.
32
The focus button for monopolar
receptacle was pressed when switch-
ing on the unit.
Check whether the correspond-
ing button was pressed when
switching on the unit.
If not, notify Erbe Service.
33
More than one button was pressed
when switching on the unit.
Check whether buttons were
pressed when switching on the
unit.
If not, notify Erbe Service.
34 – 42
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
43
The unit temperature is outside the
operational range.
If the temperature is too high,
allow the unit to cool down.
If the temperature is too low,
allow the unit to acclimate to
room temperature.
If the error still occurs, notify
Erbe Service.
44 – 50
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
Error
code
Description
Additional informa-
tion
Action
Summary of Contents for VIO 100 C
Page 1: ...10 15 80116 221 SERVICE MANUAL VIO 100 C VIO 50 C V 1 0 x V 1 0 x ELECTROSURGERY...
Page 2: ......
Page 3: ...SERVICE MANUAL VIO 100 C VIO 50 C...
Page 6: ...Table of Contents 6 42 80116 221 10 15...
Page 20: ...4 Service mode 20 42 80116 221 10 15...
Page 26: ...5 Remedying malfunctions 26 42 80116 221 10 15...