6
•
Maintenance and servicing
30
/ 42
801
16-
221
10.
15
User manual and visual inspections
•
Test specimen and accessories (where enclosed) undamaged exter-
nally.
•
User manual present.
•
All labels on the test specimen (conformity declaration mark, rating
plate, and all wording) present and readily legible.
Tests to be conducted in accordance with the national
specifications and regulations
Grounded conductor test
•
Ground terminal to chassis.
•
Ground terminal to potential equalization pin.
Ground leakage current
measurement
•
Ground leakage current, normal condition (N.C.).
•
Ground leakage current, single-fault condition (S.F.C.).
Patient leakage current
measurement
•
Patient leakage current, normal condition (N.C.).
•
Patient leakage current, single-fault condition (S.F.C.).
IMPORTANT!
Close the output relay of the unit for testing the patient leak-
age current. Procedure:
Only for V 1.0.0:
1. Call up service level (see page 19).
2. Use the Plus/Minus button to select and the yellow selection button
to activate the test program <Setup 24>.
3. Close the relay of the bipolar receptacle. Use the Plus/Minus buttons
next to the CUT and COAG displays and perform the following set-
tings:
– CUT display: <Setup 1> (= bipolar receptacle)
– COAG display:
Bipolar relay closed / monopolar relay
open
4. Close the relay of the monopolar receptacle. Use the Plus/Minus but-
tons next to the CUT and COAG displays and perform the following
settings:
– CUT display: <Setup 2> (= monopolar receptacle)
– COAG display:
Bipolar and monopolar relay closed
5. Perform the measurement.
6. Switch the test specimen off.
Summary of Contents for VIO 100 C
Page 1: ...10 15 80116 221 SERVICE MANUAL VIO 100 C VIO 50 C V 1 0 x V 1 0 x ELECTROSURGERY...
Page 2: ......
Page 3: ...SERVICE MANUAL VIO 100 C VIO 50 C...
Page 6: ...Table of Contents 6 42 80116 221 10 15...
Page 20: ...4 Service mode 20 42 80116 221 10 15...
Page 26: ...5 Remedying malfunctions 26 42 80116 221 10 15...