48
Entek IRD 6652 DSP Protection Monitor
Specifications
Power supply
Monitor
AC: 85–264 volts (50-400 Hz)
DC: 24 VDC (nominal), and must be b18 to +30 volts
Power consumption
Total module power is 30 VA
Peak load is 3.7A@24V for 250 ms following power up
Transducer
6692 unit transducer supplies: -24 VDC, +24 VDC, or -12 VDC
at 100 mA for non-contact eddy current probes
Heat production
24 VDC power: 21-22 W typical, 24 W maximum
117 VAC power: 25-26 W typical, 28 W maximum
230 VAC power: 28-29 W typical, 31 W maximum
Physical attributes
Dimensions
15 inches long, 5.75 inches high, 2.44 inches wide
27.9 cm long, 14.6 cm high, 6.19 cm wide
Weight
3.20 lbs (1.45 kg) for monitor, power supply, and backplane
Environmental
Operating temperature
Normal operating range: -4° F to +131° F (-20° C to +55° C)
Storage temperature
Storage: -4° F to +167° F (-20° C to +75° C)
Relative humidity
0 to 95%, noncondensing
Circuit board and components are conformal coated