Operating Instructions
J22 TDLAS Gas Analyzer
92
Hauser
The user is in the diagnostic message.
1.
Press
(
symbol)
The Diagnostic list submenu opens.
2.
Select the desired diagnostic event with
or
and press
.
The message for the remedial measures for the selected diagnostic event opens.
3.
Press
+
simultaneously.
The message for the remedial measures closes.
The user is in the
Diagnostics
menu at an entry for a diagnostics event, e.g., in the
Diagnostic list
submenu or
Previous diagnostics
parameter.
1.
Press
.
The message for the remedial measures for the selected diagnostic event opens.
2.
Press
+
simultaneously.
The message for the remedial measures closes.
9.3
Diagnostic information in the web browser
9.3.1
Diagnostic options
Any faults detected by the measuring device are displayed in the web browser on the home page once the user has
logged on.
A0031056-SSI
Fig 66. Diagnostic information in the web browser
1
Status area with Status signal
2
3
Remedy information with Service ID
In addition, diagnostic events which have occurred can be shown in the Diagnostics menu:
•
From parameters
•
Status signals
The Status signals are categorized in accordance with VDI/VDE 2650 and NAMUR Recommendation NE 107.
Symbol
Meaning
Failure.
A device error has occurred. The measured value is no longer valid.
Function check.
The device is in service mode (e.g., during a simulation).
1
2