The "safety settings" menu
69
4.6
"reaction high temperature"
4.6.1
"overtemperature sensor N" (N = 1 or 2)
Defines how the instrument reacts if the maximum temperature of the sensor is exceeded.
Selection:
•
warning (default)
If the maximum temperature is exceeded, an error message (A01661 or A02661) is generated
but the measurement is continued.
•
alarm
If the maximum sensor temperture is exceeded, the output assumes a defined value ("output on
alarm", see above). Additionally an error meassage (A01661 or A02661) is generated.
4.6.2
"maximum temperature sensor N" (N = 1 or 2)
Displays the maximum temperature of the respective sensor.
4.7
"defective temperature sensor"
4.7.1
"defective temperature sensor N" (N = 1 or 2)
Defines how the instrument reacts in the case of a defective temperature sensor.
Selection:
•
warning
If the temperature sensor is defective, an error message (A01661 or A02661) is generated but the
measurement is continued.
•
alarm (default)
If the temperature sensor is defective, the output assumes a defined value ("output on alarm", see
above). Additionally an error meassage (A01661 or A02661) is generated.
⇒
react. high temp
AX107
overtemp sen 1:
max. temp. Sen. 1:
overtemp sen 2:
max. temp. Sen. 2:
⇒
def. temp. sensor
AX018
def. temp. sens 1
def. temp. sens 2