5.4
Degree of protection
The measuring devices fulfill all the requirements for IP 67 (NEMA 4X).
Compliance with the following points is mandatory following installation in the field or ser-
vicing, in order to ensure that IP 67 (NEMA 4X) protection is maintained:
• The housing seals must be clean and undamaged when inserted into their grooves. The
seals must be dried, cleaned or replaced if necessary.
• All threaded fasteners and screw covers must be firmly tightened.
• The cables used for connection must be of the specified outside diameter (cable entry
• Firmly tighten the cable entries (a).
• The cables must loop down before they enter the cable entries ("water trap") (b). This
arrangement prevents moisture penetrating the entry. Always install the measuring
device in such a way that the cable entries do not point up.
• Remove all unused cable entries and insert blanking plugs instead.
• Do not remove the grommet from the cable entry.
A0001914
Fig. 25:
Installation instructions, cable entries