Proline t-mass 65 PROFIBUS DP/PA
Trouble-shooting
Hauser
105
Wall-mount housing
#
Warning!
• Risk of electric shock. Exposed components carry dangerous voltages. Switch off the power
supply before you remove the cover of the electronics compartment.
• Risk of damaging electronic components (ESD protection). Static electricity can damage
electronic components or impair their operability. Use a workplace with a grounded work-
ing surface purposely built for electrostatically sensitive devices!
• If you cannot guarantee that the dielectric strength of the measuring device can be main-
tained during the following steps, then an appropriate inspection test must be carried out
in accordance with the manufacturer’s specifications.
• When connecting Ex-certified measuring devices, see the notes and
diagrams in the Ex-specific supplement to these Operating Instructions. Please do not hes-
itate to contact your Hauser sales center if you have any questions.
"
Caution!
Use only original Hauser parts.
Removal and installation
1.
Remove the screws and open the hinged cover (1) of the housing.
2.
Loosen the screws securing the electronics module (2). Then push up electronics mod-
ule and pull it as far as possible out of the wall-mount housing.
3.
Disconnect the sensor signal cable plug including HistoROM/S-DAT (7.2) and Histo-
ROM/T-DAT (7.3) from amplifier board (7.1).
4.
Remove the cover (4) from the electronics compartment by removing the screws.
5.
Disconnect the ribbon cable plug (3) of the display module from amplifier board.
6.
Remove the boards (6, 7, 8):
Insert a thin pin into the hole (5) provided for the purpose and pull the board clear of
its holder.
7.
Installation is the reverse of the removal procedure.