Trouble-shooting
Proline t-mass 65 PROFIBUS DP/PA
96
Hauser
032
S: SENSOR SW DAT
$
: # 032
0x10
0x11
0x12
BAD
Sensor fail-
ure
O.K.
Low
High
S-DAT data
inconsistent
Cause:
Error accessing the calibration values stored in
the HistoROM/S-DAT.
Remedy:
1. Check whether the HistoROM/S-DAT is
correctly plugged into the amplifier board
103,
2. Replace the HistoROM/S-DAT if it is defec-
tive.
Before replacing the DAT, check that the
new, replacement DAT is compatible with
the measuring electronics. Check the:
– Spare parts set number
– Hardware revision code
3. Replace measuring electronics boards if
necessary.
035
S: SENS HW-ROM/
RAM
$
: # 035
0x10
0x11
0x12
BAD
Sensor fail-
ure
O.K.
Low
High
Sensor:
Defective ROM/
RAM
Cause:
Sensor: Defective ROM/RAM
Remedy:
Replace the remote amplifier board.
036
S: SENS SW-ROM/RAM
$
: # 036
0x10
0x11
0x12
BAD
Sensor fail-
ure
O.K.
Low
High
Sensor:
Defective ROM/
RAM
Cause:
Sensor: Defective ROM/RAM
Remedy:
Replace the amplifier board.
042
S: TRANSM. SW-DAT
$
: # 042
0x0F
BAD
Device Fail-
ure
Constant
T-DAT data
inconsistent
Cause:
Error accessing the calibration values stored in
the HistoROM/T-DAT.
Remedy:
1. Check whether the HistoROM/T-DAT is
correctly plugged into the amplifier board.
2. Replace the HistoROM/T-DAT if it is defec-
tive.
102,
Before replacing the DAT, check that the
new, replacement DAT is compatible with
the measuring electronics. Check the:
– Spare parts set number
– Hardware revision code
3. Replace measuring electronics boards if
necessary.
051
AC COMPATIB.
$
: # 051
0x0F
BAD
Device Fail-
ure
Constant
Compatibility
Amp. I/O module
Cause:
The I/O board and the amplifier board are not
compatible.
Remedy:
Use only compatible modules and boards.
Check the compatibility of the modules used.
Check the:
– Spare parts set number
– Hardware revision code
No.
Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic mes-
sage in the PRO-
FIBUS
master
Cause/
Rectification
Spare parts
→
102
Q
ual
ity co
de (HEX
)
M
ea
sured v
aria
b
le status
Qu
ality
sta
tus
Q
ual
ity
substa
tu
s
Li
mits
Summary of Contents for Proline t-mass 65 PROFIBUS DA