Proline t-mass 65 PROFIBUS DP/PA
Trouble-shooting
Hauser
95
10.2.2
List of system error messages
No.
Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic mes-
sage in the PRO-
FIBUS
master
Cause/
Rectification
Spare parts
Qu
ality
cod
e (HE
X
)
Measu
red
v
ariab
le sta
tus
Qu
al
it
y stat
us
Quality substatus
Limits
S = System error
$
= Fault message (with an effect on the current operation)
! = Notice message (without any effect on the current operation)
No. # 0xx
Hardware error
001
S: CRITICAL FAIL
$
: # 001
0x0F
BAD
Device Fail-
ure
Constant
ROM/RAM
failure
Cause:
Serious device error
Remedy:
Replace the amplifier board.
011
S: AMP HW EEPROM
$
: # 011
0x0F
BAD
Device Fail-
ure
Constant
Amplifier
EEPROM
failure
Cause:
Amplifier: Defective EEPROM
Remedy:
Replace the amplifier board.
012
S: AMP SW EEPROM
$
: # 012
0x0F
BAD
Device Fail-
ure
Constant
Amp. EEPROM
data inconsistent
Cause:
Measuring amplifier: Error when accessing data
of the EEPROM
Remedy:
Replace the amplifier board.
014
S: AMP SW-ROM/RAM
$
: # 014
0x0F
BAD
Device Fail-
ure
Constant
Amp. ROM/RAM
data inconsistent
Cause:
Amplifier: Defective ROM/RAM
Remedy:
Replace the amplifier board.
031
S: SENSOR HW DAT
$
: # 031
0x10
0x11
0x12
BAD
Sensor fail-
ure
O.K.
Low
High
S-DAT failure /
not inserted
Cause:
Sensor DAT:
1. HistoROM/S-DAT is defective.
2. HistoROM/S-DAT is not plugged into the
amplifier board or is missing.
Remedy:
1. Replace the S-DAT.
Check the spare parts set number to ensure
that the new, replacement DAT is compati-
ble with the measuring electronics.
2. Plug the HistoROM/S-DAT into the
amplifier board
Summary of Contents for Proline t-mass 65 PROFIBUS DA