Installation
Proline t-mass 65 PROFIBUS DP/PA
20
Hauser
A0005118
Fig. 10:
Dimensions needed to calculate the insertion depth
A
Pipes: internal diameter
Ducts: internal dimension
B
Wall thickness
C
Dimension from pipe/duct to the compression fitting
The following dimensions are required to calculate the insertion depth:
!
Note!
For detailed remarks on calculation refer to Technical Information TI00069D.
• Calculated insertion depth = (0.3 × A) + B + C + 2 mm (0.08 in)
Note down the calculated value.
A0010001
Fig. 11:
Aligning the sensor to the calculated insertion depth
A
B
C
A
B
C
230
220
210
200
190
180
9
8
7
230
220
210
200
190
180
9
8
7
230
220
210
200
190
180
9
8
7
A
•
For circular pipes: the internal diameter (DN)
•
For rectangular ducts:
– The internal duct height if the sensor is installed vertically
– The internal duct width if the sensor is installed horizontally
!
Note!
Minimum length of dimension A = 80 mm (3.15 in)
B
Pipe / duct wall thickness
C
Height of the welding nozzle at the pipe/duct including the sensor compression fitting or low pressure
mounting set (if used).
1
3
4
2
230
220
210
200
190
180
9
8
7