Proline Prowirl 72
5 Operation
E Hauser
37
5.4.4
Device status / error messages
You can read the extended device status, in this case, current error messages, via command “48”.
The command delivers bit-encoded information (see table below).
!
Note!
Detailed information on the device status messages and error messages, and how they are rectified,
can be found on Page 48 ff.!
Byte
Bit
Error no.
Short error description (
→
Page 48 ff.)
0
0
001
Serious device error.
1
011
Faulty amplifier EEPROM.
2
012
Error when accessing data of the amplifier EEPROM.
3
021
COM module:
Faulty EEPROM.
4
022
COM module:
Error when accessing EEPROM data.
5
111
Totalizer checksum error.
6
351
Current output: the current flow is outside the set range.
7
Not assigned
–
1
0
359
Pulse output: the pulse output frequency is outside the set range.
1
Not assigned
–
2
379
Device being operated in its resonance frequency.
3
Not assigned
–
4
Not assigned
–
5
394
DSC sensor defective, no measurement.
6
395
DSC sensor being operated near application limits, device failure probable
soon.
7
396
Device finds signal outside the set filter range.
2
0...1
Not assigned
–
2
399
Pre-amplifier disconnected.
3...5
Not assigned
–
6
501
Loading a new amplifier software version or data into the device. No other
commands possible at this point.
7
502
Uploading the device data.
No other commands possible at this point.
3
0
601
Positive zero return active.
1
611
Current output simulation active.
2
Not assigned
–
3
631
Pulse output simulation active.
4
641
Status output simulation active.
5
691
Simulation of failsafe mode (outputs) active.
6
692
Simulation measurand.
7
Not assigned
–
4
0...1
Not assigned
–
2
698
Current adjustment active
3...7
Not assigned
–
Summary of Contents for Proline Prowirl 72
Page 4: ...QUICK SETUP for quick commissioning Proline Prowirl 72 4 Endress Hauser...
Page 20: ...3 Installation Proline Prowirl 72 20 Endress Hauser...
Page 42: ...6 Commissioning Proline Prowirl 72 42 Endress Hauser...
Page 44: ...7 Maintenance Proline Prowirl 72 44 Endress Hauser...
Page 114: ...www endress com worldwide BA084D 06 en 12 05 71008404 FM SGML 6 0...