12 Factory settings
Proline Prowirl 72
106
E Hauser
Switch-on point and switch-off point, Prowirl W (see Page 90)
The factory settings in the table are given in the unit dm³/s. If another unit is selected in the UNIT
FLOW function (see Page 74), the corresponding value is converted and displayed in the selected
unit.
Switch-on point and switch-off point, Prowirl F (see Page 90)
The factory settings in the table are given in the unit dm³/s. If another unit is selected in the UNIT
FLOW function (see Page 74), the corresponding value is converted and displayed in the selected
unit.
Nominal diameter
DN
Gas
Liquid
DIN
[mm]
ANSI
[inch]
Switch-on point
[dm³/s]
Switch-off point
[dm³/s]
Switch-on point
[dm³/s]
Switch-off point
[dm³/s]
15
½"
13
10
2.1
1.7
25
1"
49
40
5.9
4.8
40
1½"
110
94
14
11
50
2"
190
150
22
18
80
3"
420
340
50
41
100
4"
710
580
85
70
150
6"
1600
1300
190
160
200
8"
–
–
–
–
250
10"
–
–
–
–
300
12"
–
–
–
–
Nominal diameter
DN
Gas
Liquid
DIN
[mm]
ANSI
[inch]
Switch-on point
[dm³/s]
Switch-off point
[dm³/s]
Switch-on point
[dm³/s]
Switch-off point
[dm³/s]
15
½"
7.7
6.3
1.5
1.2
25
1"
38
31
4.6
3.8
40
1½"
94
77
11
9.2
50
2"
160
130
19
15
80
3"
350
290
42
35
100
4"
610
500
73
60
150
6"
1400
1100
170
140
200
8"
2700
2200
320
260
250
10"
4200
3400
500
410
300
12"
6000
4900
720
590
Summary of Contents for Proline Prowirl 72
Page 4: ...QUICK SETUP for quick commissioning Proline Prowirl 72 4 Endress Hauser...
Page 20: ...3 Installation Proline Prowirl 72 20 Endress Hauser...
Page 42: ...6 Commissioning Proline Prowirl 72 42 Endress Hauser...
Page 44: ...7 Maintenance Proline Prowirl 72 44 Endress Hauser...
Page 114: ...www endress com worldwide BA084D 06 en 12 05 71008404 FM SGML 6 0...