Error message display
Type of error
Errors which occur during commissioning or measuring operation are displayed immediately. If two
or more system or process errors occur, the error with the highest priority is always the one shown
on the display. The measuring system distinguishes between two types of error:
•
System error:
this group includes all device errors, for example communication errors, hardware
errors, etc.
→
Page 48
•
Process error:
this group includes all application errors, for example “DSC SENSOR LIMIT”, etc.
→
Page 48
A0000991
Fig. 23:
Error messages on the display (example)
1
Type of error: P = Process error, S = System error
2
Error message type:
$
= Fault message, ! = Notice message (definition: see below)
3
Error designation: e.g. DSC SENS LIMIT = Device being operated near application limits
4
Error number: e.g. #395
5
Duration of most recent error occurrence (in hours, minutes and seconds), display format - see OPERATION
HOURS function on Page 103
Type of error message
Users have the option of weighting system and process errors differently by defining them as
Fault
messages
or
Notice messages
. This is specified via the function matrix (
→
see SUPERVISION
function group on Page 102).
Serious system errors, e.g. electronic module defects, are always categorized and displayed as “fault
messages” by the measuring device.
Notice message (!)
• Displayed as
→
exclamation mark (!), error group (S: system error, P: process error).
• The error in question has no effect on the inputs or outputs of the measuring device.
Fault message (
$
)
• Displayed as
→
lightning flash
(
$
)
, error designation (S: system error, P: process error)
• The error in question has a direct effect on the inputs or outputs.
The response of the inputs/outputs (failsafe mode) can be defined by means of functions in the
function matrix (see Page 52).
!
Note!
Error messages can be output via the current output in accordance with NAMUR NE 43.
1
2
4
5
3
X X X X X X X X X X
# 0 0 0 0 0 : 0 0 : 0 5
P
Summary of Contents for Proline Prowirl 72
Page 4: ...QUICK SETUP for quick commissioning Proline Prowirl 72 4 Endress Hauser...
Page 20: ...3 Installation Proline Prowirl 72 20 Endress Hauser...
Page 42: ...6 Commissioning Proline Prowirl 72 42 Endress Hauser...
Page 44: ...7 Maintenance Proline Prowirl 72 44 Endress Hauser...
Page 114: ...www endress com worldwide BA084D 06 en 12 05 71008404 FM SGML 6 0...