Proline Promass O 100 HART
Technical data
Hauser
103
Calculation of the maximum measured error as a function of the flow rate
Flow rate
Maximum measured error in % o.r.
ZeroPoint
BaseAccu
⋅
100
³
A0021332
± BaseAccu
A0021339
ZeroPoint
BaseAccu
⋅
100
<
A0021333
ZeroPoint
MeasValue
⋅
100
±
A0021334
Calculation of the maximum repeatability as a function of the flow rate
Flow rate
Maximum repeatability in % o.r.
½
⋅
ZeroPoint
BaseRepeat
⋅
100
³
A0021335
A0021340
½
⋅
ZeroPoint
BaseRepeat
⋅
100
<
A0021336
ZeroPoint
MeasValue
⋅
100
½
⋅
±
A0021337
Example for max. measured error
E [%]
Q [%]
0
5
10
15
20
25
30
35
40
45
50
60
65
55
0
0.5
1.0
1.5
2.0
2.5
70
75
80
85
90
95
100
A0016708
E
Error: Maximum measured error as % o.r. (example using PremiumCal)
Q
Flow rate as %
16.7 Installation
"Mounting requirements" (→ 18)
16.8 Environment
Ambient temperature
range
Storage temperature
–50 to +60 °C (–58 to +140 °F) (Order code for "Test, certificate", option JM)
Climate class
DIN EN 60068-2-38 (test Z/AD)
Summary of Contents for Proline Promass O 100HART