Diagnostics and troubleshooting
Proline Promass O 100 HART
82
Hauser
In the case of some items of diagnostic information, the status signal and the
diagnostic behavior can be changed. Adapt the diagnostic information (→ 81)
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
Diagnostic of sensor
022
Sensor temperature
1.Change main electronic module
2.Change sensor
F
Alarm
046
Sensor limit exceeded
1. Inspect sensor
2. Check process condition
S
Alarm
062
Sensor connection
1.Change main electronic module
2.Change sensor
F
Alarm
082
Data storage
1. Check module connections
2. Contact service
F
Alarm
083
Memory content
1. Restart device
2. Contact service
F
Alarm
140
Sensor signal
1.Check or change main
electronics
2.Change sensor
S
Alarm
144
Measuring error too high
1. Check or change sensor
2. Check process conditions
F
Alarm
190
Special event 1
Contact service
F
Alarm
191
Special event 5
Contact service
F
Alarm
192
Special event 9
Contact service
F
Alarm
Diagnostic of electronic
201
Device failure
1. Restart device
2. Contact service
F
Alarm
242
Software incompatible
1. Check software
2. Flash or change main
electronics module
F
Alarm
252
Modules incompatible
1. Check electronic modules 2.
Change electronic modules
F
Alarm
262
Module connection
1. Check module connections
2. Change main electronics
F
Alarm
270
Main electronic failure
Change main electronic module
F
Alarm
271
Main electronic failure
1. Restart device
2. Change main electronic module
F
Alarm
272
Main electronic failure
1. Restart device
2. Contact service
F
Alarm
273
Main electronic failure
Change electronic
F
Alarm
274
Main electronic failure
Change electronic
S
Warning
283
Memory content
1. Reset device 2. Contact service
F
Alarm
302
Device verification active
Device verification active, please
wait.
C
Warning
311
Electronic failure
1. Reset device 2. Contact service
F
Alarm
311
Electronic failure
1. Do not reset device
2. Contact service
M
Warning
375
I/O communication failed 1. Restart device
2. Change main electronic module
F
Alarm
382
Data storage
1. Insert DAT module 2. Change
DAT module
F
Alarm