Proline Promass O 100 HART
Technical data
Hauser
99
16.5 Power supply
Terminal assignment
Pin assignment, device plug
Supply voltage
Transmitter
For device version with all communication types except Modbus RS485 intrinsically safe:
DC 20 to 30 V
The power unit must be tested to ensure it meets safety requirements (e.g. PELV, SELV).
Power consumption
Transmitter
Order code for
"Output"
Maximum
Power consumption
Option
B
: 4-20mA HART, pulse/frequency/switch output
3.5 W
Current consumption
Transmitter
Order code for
"Output"
Maximum
Current consumption
Maximum
switch-on current
Option
B
: 4-20mA HART, pul./
freq./switch output
145 mA
18 A (<0.125 ms)
Power supply failure
• Totalizers stop at the last value measured.
• Depending on the device version, the configuration is retained in the device memory or
in the plug-in memory (HistoROM DAT).
• Error messages (incl. total operated hours) are stored.
Electrical connection
Potential equalization
No special measures for potential equalization are required.
Terminals
Transmitter
Spring terminals for wire cross-sections0.5 to 2.5 mm
2
(20 to 14 AWG)
Cable entries
• Cable gland: M20 × 1.5 with cable 6 to 12 mm (0.24 to 0.47 in)
• Thread for cable entry:
– NPT ½"
– G ½"
– M20
Cable specification
Summary of Contents for Proline Promass O 100HART