Proline Promass O 100 HART
Technical data
Hauser
95
16
Technical data
16.1 Application
The measuring device is suitable for flow measurement of liquids and gases only.
Depending on the version ordered, the measuring device can also measure potentially
explosive, flammable, poisonous and oxidizing media.
To ensure that the device remains in proper operating condition for its service life, use the
measuring device only for media against which the process-wetted materials are
adequately resistant.
16.2 Function and system design
Measuring principle
Mass flow measurement based on the Coriolis measuring principle
Measuring system
One device version is available: compact version, transmitter and sensor form a mechanical
unit.
For information on the structure of the device (→ 11)
16.3 Input
Measured variable
Direct measured variables
• Mass flow
• Density
• Temperature
Calculated measured variables
• Volume flow
• Corrected volume flow
• Reference density
Measuring range
Measuring ranges for liquids
DN
Measuring range full scale values
min(F)
to
max(F)
[mm]
[in]
[kg/h]
[lb/min]
80
3
0 to 180
000
0 to 6
615
100
4
0 to 350
000
0 to 12
860
150
6
0 to 800
000
0 to 29
400
Measuring ranges for gases
The full scale values depend on the density of the gas and can be calculated with the
formula below:
max(G)
=
max(F)
· ρ
G
: x
max(G)
Maximum full scale value for gas [kg/h]
max(F)
Maximum full scale value for liquid [kg/h]