Commissioning
Proline Promag L 400 HART
88
Hauser
Navigation path
"Setup" menu
®
Advanced setup
®
Sensor adjustment
Structure of the submenu
Sensor adjustment
®
Installation direction
Parameter overview with brief description
Parameter
Description
Selection/
Factory setting
Installation direction
Change the sign of the
direction of flow of the fluid.
• Flow in arrow direction
• Flow against arrow
direction
Flow in arrow direction
10.5.4 Performing electrode cleaning
The parameters of the electrode cleaning circuit can be configured in the
Electrode cleaning circuit
(ECC)
submenu.
Navigation path
"Setup" menu
®
Advanced setup
®
Electrode cleaning circuit
The wizard only appears if the device was ordered with an electrode cleaning circuit.
Structure of the submenu
Electrode cleaning circuit
®
Electrode cleaning circuit
ECC duration
ECC recovery time
ECC cleaning cycle
ECC polarity
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection/
User entry
Factory setting
Electrode cleaning
circuit
Enable the cyclic
electrode cleaning
circuit.
• Off
• On
Off
ECC duration
The following option is
selected in the
Electrode cleaning
circuit
parameter:
On
Specify the cleaning
duration of the
electrode cleaning
circuit.
1 to 30 s
2 s