10.4.7 Configuring empty pipe detection
The
Empty pipe detection (EPD)
wizard guides you systematically through all parameters that
have to be set for configuring empty pipe detection.
Navigation path
"Setup" menu
®
Empty pipe detection
Sequence of the wizard
Empty pipe det.
Response time
End of wizard
New adjustment
Cancel
Empty pipe adj.
Progress
Full pipe adjust.
Ok
Busy
Not ok
On
Switch point EPD
Off
A0017210-EN
å
30
"Empty pipe detection" wizard in the "Setup" menu
Parameter overview with brief description
Parameter
Description
Selection/
User entry
Factory setting
Empty pipe detection
Switch empty pipe detection on
and off.
• Off
• On
Off
New adjustment
Select the type of adjustment.
• Cancel
• Empty pipe adjustment
• Full pipe adjustment
Cancel
Progress
Select the progress display.
• Ok
• Busy
• Not OK
Not OK
Switch point empty pipe
detection
Enter the switch point for empty
pipe detection.
1 to 99 %
50 %