Liquisys M CPM223/253
Trouble-shooting
Hauser
93
Additional functions
(S-package) missing
No or incorrect release code used
If retrofitting: check whether the correct
serial number was quoted when ordering
the S-package.
Handled by Hauser Sales
Incorrect device serial number saved in
LSCH/LSCP module
Check whether serial number on the
nameplate matches SNR in LSCH/ LSCP
(field S 8).
The serial number of the device is definitive
for the S-package.
Additional functions
(S-package and/or
Chemoclean) missing after
LSCH/LSCP module
replaced
Replacement modules LSCH or LSCP
have the
device
serial number 0000
when they leave the factory. The
S-package or Chemoclean are not
enabled on leaving the factory.
In the case of LSCH/LSCP with SNR 0000,
a
device
serial number can be entered
once in fields E114 to E116.
Then enter the release code for the
S-package and/or Chemoclean.
For a detailed description, see "Replacing
central module" section.
No HART or
PROFIBUS-PA/-DP
interface function
Incorrect central module
HART: LSCH-H1 or H2 module,
PROFIBUS-PA: LSCP-PA module,
PROFIBUS-DP: LSCP-DP module,
see field E112.
Replace central module;
User or Hauser Service
Wrong software
SW version see field E111.
SW can be modified with optoscope.
Bus problem
Remove some devices and repeat the test.
Contact Hauser Service.
Error
Possible cause
Tests and/or remedial measures
Execution, tools, spare parts