Liquisys M CPM223/253
Wiring
Hauser
25
4.2.2
Measuring cable and sensor connection
To connect pH electrodes with MemoSens functionality to the transmitter, you require the
terminated data transmission cable CYK10 with 2x2 cores, twisted pair, screen and PVC jacket.
Structure of the measuring cable
C07-CPM2x3xx-04-06-00-xx-003.eps
Fig. 24:
Structure of the CYK10 measuring cable
1
Coupling (to connect to the sensor) with integrated electronics
!
Note!
For further information on the cable, please refer to the "Accessories" section.
To plug the cable coupling onto the sensor plug-in head, proceed as follows:
YE
GN
BN
WH
+ 87
- 88
Com A 96
Com B 97
1
C07-MemoSnxx-11-05-00-xx-002.eps
Fig. 25:
Handling of sensor plug-in head and cable
coupling
1. Rotate the lower part of the coupling in such
a way that the two pairs of keys (pos. 1, 2)
are located above each other.
2. Plug the coupling onto the plug-in head so
that the keys engage in the slots of the
plug-in head (pos. 3).
3. Turn the lower part of the coupling (pos. 4)
clockwise as far as possible (approx. 60°).
Doing so locks the coupling and prevents the
connection from opening inadvertently.
Open the connection in the reverse sequence of
operations.
60°
1
2
3
4