Operation
Liquisys M CPM223/253
42
Hauser
Current output 1
Current output 2
Value pair
pH / mV /
% / °C
Current
[mA]
Distance
per mA
pH / mV /
% / °C
Current
[mA]
Distance
per mA
1
2
3
4
5
6
7
8
9
Coding
Field
Setting range
(Factory settings, bold)
Display
Info
O
CURRENT OUTPUT
function group
Configuration of the current output (does not
apply for PROFIBUS).
O1
Select current output
Out1
Out 2
A characteristic can be selected for every
output.
O2
Select measured
variable for 2nd
current output
°C
pH
mV
Contr
R237/R 266 = curr (current output 2) can
only be selected if O2 = Contr is selected
(relay board required).
O3
O3 (1)
Enter or output linear
characteristic
Lin = linear
(1)
Sim = simulation (2)
Tab = table (3)
The characteristic can have a positive or
negative slope for the measured value
output.
In the case of actuating variable output
(O2 = Contr), an increasing current
corresponds to an increasing actuating
variable.
O311
Select current range
4 ... 20 mA
0 ... 20 mA
O
OUTPUT
O1
Sel.Out
Out1
O2
Parameter
°C
O3
Sel.Type
lin
O311
Sel.Range
4-20