Liquisys M CPM223/253
Operation
Hauser
63
5.4.8
E+H Service
S9
Reset the device to the
basic settings
No
Sens = sensor data
Facty = factory settings
Sens = last calibration is deleted and is reset to
factory setting.
Facty = all data (apart from A1 a. S1) are
deleted and reset to the factory setting!
S10
Perform device test
No
Displ = display test
S11
Reference voltage is
displayed
Current value in mV
This is used to check the reference potential.
Value > 50 mV indicates galvanic voltage in
the medium.
High values (> 1000 mV) may falsify the
measured value.
S12
Select AC frequency
50 Hz
60 Hz
Only select 60 Hz if the frequency of the
voltage at the place of use is 60 Hz, the
measured value is fluctuating or sporadic SCS
errors are reported.
Coding
Field
Setting range
(Factory settings, bold)
Display
Info
S9
S.Default
no
S10
Test
no
S11
mV
uncompRef
25
S12
Hz
AC Freq
50
Coding
Field
Setting range
(Factory settings, bold)
Display
Note
E
E+H SERVICE
function group
Information on the device version
E1
Select module
Contr = controller
(1)
Trans = transmitter (2)
Main = power unit (3)
Rel = relay module (4)
Sens = sensor (5)
The "Sens = sensor" option is only available
on devices with Memosens functionality.
E111
E121
E131
E141
E151
Software version is
displayed
If E1 = contr: instrument software
If E1 = trans, main, rel: module firmware
If E1 = sens: sensor software
E112
E122
E132
E142
E152
Hardware version is
displayed
Only display function
E113
E123
E133
E143
E153
Serial number is
displayed
Only display function
E
E+H SERV
E1
Select
Contr
E111
SW-Vers.
xx.xx
E112
HW-Vers.
xx.xx
E113
12345678
SerNo