Liquisys M CPM223/253
Wiring
Hauser
21
Examples for connecting pH and redox sensors
The following diagrams show the connection of various pH and redox sensors.
C07-CPM2x3xx-04-06-00-xx-010.eps
Fig. 19:
Connecting glass electrode CPS11 with CPK9 (left) and ISFET sensor CPS471 with CPK12 (right) to Liquisys M
A
Panel-mounted instrument
B
Field instrument
C
Potential matching PM for symmetrical connection
8. Push the BNC connector housing 9 over the
cable. The inner conductor must be located at
clamping surface 10 of the connector.
9. Tighten the cable gland 5.
10. Insert the clamp element 11 and screw in the
connector cover 12. This creates a safe
connection between the inner conductor and
the connector pin.
C07-CPM223xx-04-06-00-xx-016.eps
Fig. 18:
Mounting the pH connecting cable in the BNC
elbow plug
5
9
10
11
12
J
S
S
DRN
PA/
PM
PA/
PM
CPK9
CPK12
CPS11
CPS471
B
C
C
B
S
GN WH YE BN
Schir
m
/
Scr
een
11 12 13
11 12 13
BN
BN
BNC
BK
BK
BK
innen /
internal
innen /
internal
SRC
Ref
SRC
Ref
Ref
pH/
Redox
GN WH YE RD BN
PE
PE
A
A
)