Trouble-shooting
Liquisys M CPM223/253
86
Hauser
9
Trouble-shooting
9.1
Trouble-shooting instructions
The transmitter constantly monitors its functions itself. If an error occurs which the device
recognises, this is indicated on the display. The error number is under the unit display of the main
measured value. If more than one error occurs, you can call these up with the MINUS key.
Refer to the "System error messages" table for the possible error numbers and remedial measures.
Should a malfunction occur without any transmitter error message, please refer to the
"Process-specific errors" or the "Device-specific errors" tables to localise and rectify the error. These
tables provide you with additional information on any spare parts required.
9.2
System error messages
You can display and select the error messages with the MINUS key.
Error
no.
Display
Tests and/or remedial measures
Alarm contact
Error current
Autom. cleaning
start
Facty
User
Facty
User
Facty
User
E001
EEPROM memory error
1.
Switch device off and then on again.
2.
Load device software compatible with the
hardware (with optoscope, see "Optoscope
service tool" section).
3.
Load measurement-parameter specific device
software.
4.
If the error persists, send in the device for repair
to your local Hauser subsidiary or
replace the device.
Yes
No
—
—
1
E002
Instrument not calibrated, calibration
data invalid, no user data invalid
(EEPROM error), instrument
software not suitable to hardware
(controller)
Yes
No
—
—
1
E003
Download error
Invalid configuration. Repeat download, check
optoscope.
Yes
No
No
E004
Instrument software version not
compatible with module hardware
version
Load software compatible with hardware
Load measurement-parameter specific device
software.
Yes
No
No
E007
Transmitter malfunction, instrument
software not compatible with
transmitter version
Yes
No
—
—
1
E008
SCS alarm:
Glass electrode: glass breakage
ISFET: leakage current > 400 nA
Check glass electrode for glass breakage or hair-line
cracks;
Inspect electrode plug-in head for moisture and dry
if necessary;
Check medium temperature.
Replace ISFET.
Yes
No
No
E010
Temperature sensor defective, not
connected or short-circuited
Check temperature sensor and connections; check
device and measuring cable with temperature
simulator if necessary.
Check correct option selected in field A5.
Yes
No
No
E030
SCS reference electrode warning
Check reference electrode for contamination and
damage; clean electrode.
No
No
No
E032
Slope range exceeded or below range
Repeat calibration and renew buffer solution;
replace electrode if necessary, check device and
measuring cable with simulator.
Yes
No
—
—
1
E033
pH value zero too low or too high
Yes
No
—
—
1
E034
Redox offset range exceeded or
below range
Yes
No
—
—
1
E041
Calculation of calibration parameters
aborted
Repeat calibration and renew buffer solution;
replace electrode if necessary, check device and
measuring cable with simulator.
Yes
No
—
—
1