Liquisys M CPM223/253
Operation
Hauser
43
O312
0/4 mA value:
Enter associated pH
(redox) or temperature
value
pH 2.00
pH -2.00 ... 16.00
-1500 mV
-1500 ... 1500 mV
0.0 %
0.0 ...100.0 %
0.0 C
-20 ... 150.0 °C
Here you can enter the measured value at
which the min. current value (0/4 mA) is
applied at the transmitter output.
(Spreading: see Technical data.)
O313
20 mA value:
Enter associated pH
(redox) or temperature
value
pH 12.0
pH -2.00... 16.00
1500 mV
-1500 mV ... 1500 mV
100.0 %
0.0 ...100.0 %
100.0 C
-20.0 ... 150.0 °C
Here you can enter the measured value at
which the max. current value (20 mA) is
applied at the transmitter output.
(Spreading: see Technical data.)
O3 (2)
Simulate current
output
Lin = linear (1)
Sim = simulation
(2)
Tab = table (3)
Simulation is not ended until (1) or (3) is
selected.
For further characteristics, see O3 (1), O3(3).
O321
Enter simulation value
Current value
0.00 ... 22.00 mA
Entering a current value results in this value
being directly output at the current output.
O3 (3)
Enter current output
table (only for Plus
Package)
Lin = linear (1)
Sim = simulation (2)
Tab = table
(3)
Values can also be added or altered at a later
stage. The values entered are automatically
sorted by increasing current value.
For further characteristics, see O3 (1), O3
(2).
O331
Select table options
Read
Edit
O332
Enter number of table
value pairs
1
1 ... 10
Enter the number of pairs from the x and y
value (measured value and current value)
here.
O333
Select table value pair
1
1 ... No. elem.
Assign
O334
Enter x value
pH 0.00
pH -2.00 ... 16.00
0 mV
-1500 ... 1500 mV
0.0 %
0.0 ... 100.0 %
x value = measured value specified by user.
Coding
Field
Setting range
(Factory settings, bold)
Display
Info
pH
O312
0/4 mA
2.00
pH
O313
20 mA
12.00
O3
Sel.Type
sim
O321
mA
Simulat.
10.20
O3
Sel.Type
table
O331
Sel.Table
read
O332
No.Elem.
1
O333
Sel.Elem.
1
O334
pH
Meas.val.
0.00