7.2
Troubleshooting using the error messages
Display and select the error messages using the MINUS key.
Error
no.
Cause
Measures
Contact
Error current
Automatic
cleaning trigger
Fact
User
Fact
User
Fact
User
E001
EEPROM memory error
Switch instrument off and back on,
return instrument to your local
Hauser sales agency for repair or
replace instrument.
Load software compatible with hardware
Load instrument software specific to
parameter measured.
yes
no
—
—*
E002
Instrument not calibrated,
calibration data invalid,
no user data available or
user data invalid
(EEPROM error)
Instrument software not
suitable for hardware
(Controller)
yes
no
—
—*
E003
Download error
Invalid configuration. Repeat download,
check optoscope.
yes
no
—
—*
E004
Device software version
incompatible with
hardware version of
module
Load software compatible with hardware
Load instrument software specific to
parameter measured.
yes
no
—*
E007
Transmitter malfunction
Software does not match
hardware (transmitter)
yes
no
—
—*
E032
Below slope range or
range exceeded
Service sensor and recalibrate.
yes
no
—
—*
E044
Measured value unstable
(during calibration)
Service sensor and recalibrate.
yes
no
no
E057
Measuring range of main
parameter exceeded
Check measurement, control and
connections.
yes
no
no
E059
Below temperature value
measuring range
yes
no
no
E061
Temperature value
measuring range
exceeded
yes
no
no
E063
Current output 1: range
undershot
Check configuration.
yes
no
no
E064
Current output 1: range
exceeded
Check measured value and current
assignment.
yes
no
no
E065
Current output 2: range
undershot
yes
no
no
E066
Current output 2: range
exceeded
yes
no
no
E067
Alarm threshold limit
contactor 1 exceeded
Limit or reference value exceeded or
undershot.
yes
no
no
E068
Alarm threshold limit
contactor 2 exceeded
yes
no
no
E069
Alarm threshold limit
contactor 3 exceeded
yes
no
no
E070
Alarm threshold limit
contactor 4 exceeded
yes
no
no
E080
Distance between
0/4 mA and 20 mA values
at output 1 too short
Adjust spreading for current output 1.
no
no
—
—*
OM253FE7.CHP
Liquisys M COM 223 F / 253 F
Maintenance and troubleshooting
Hauser
55