Coding
Field
Selection or range
default (bold)
Display
Info
C
Function group
CALIBRATION
C1 (1)
Calibration
in air
Air
Ref
Remove the sensor from the medium.
C111
Start calibration
Last sensor slope
Counter in second
row counts down:
600 s
→
0 s
The sensor slope is checked for
10 s after 530 s (desired range
75 % ... 140 %, outside range:
error 32, calibration is aborted).
The stability of the signal is
checked during the last 60 s of the
calibration (< 1 %), if greater:
error 45 and calibration is
aborted.
C112
Calibration status
is displayed
o.k.
E xxx
Immerse sensor in medium, if C112 = o.k.
C113
Save calibration
result?
yes
no
new
If C112 = E xxx, then only no or
new.
If new, return to C.
If yes/no, return to »Measure-
ment«
C1 (2)
Single point
calibration
in fluid
Air
Ref
C121
Enter calibration
value
Current value
Current value must be
determined through reference
measurement.
The factory settings are printed in bold; functions printed in
italics are not available in the basic version.
C
CALIBRAT
Air
C1
Calibrat
100
C111
600
%
o.k.
C112
Status
yes
C113
Store
Ref
C1
Calibrat
10.00
C121
Real
PV
mg/l
Instrument configuration
Liquisys M COM 223 F / 253 F
48
Hauser