Appendix
Function group
CALIBRATION
C
MEAS. VALUE DISPLAY
with
TEMPERAT. DISPLAY
in °C
Measured value display
in mg/l
Measured value display
in % SAT
Measured value display
in hPa
Temperature display
suppressed
Temperature display
in °F
Edit:
Co
de
22
Re
ad
:
any Co
de
Edit:
Co
de
22
Re
ad
:
any Co
de
other errors are
displayed
(up to 10 errors)
1st error is displayed
(if present)
Function group
SETUP 1
A
Selection of Operation mode
%SAT
mg/l
A1
Switch pressure
compensation on or off
; on
off
Select display unit
;
ppm
mg/l
A2
Entry of Altitude
0 ... 4000 m
0 m
A3
Entry of measuring
value damping
(no damping)
1 ... 60
1
.
Selection of
measuring range
0 …
0 …
20 mg/l
200 % SAT
A4
A5
A5
Function group
SETUP 2
B
Entry salt concentration
0 ... 4.0 %
0.0 %
B1
Entry of correct
process temperature
–10.0 ... 60.0
current meas. value
°C
B2
Display of temperature
difference (Offset)
–5.0 ... 5.0 °C
current offset
B3
Calibration
Ref = entry of
reference value
Air
C1 (2)
C1 (1)
Start calibration
Countdown
600 s
0 s
→
Calibration status
is displayed
o.k.
E---
C111
C112
Store
calibration results
; no; new
yes
C113
Entry of calibration value
current value
C121
Calibration status
is displayed
o.k.
E---
C123
Slope is
displayed
75.0 ... 140 %
100.0 %
C122
Store
calibration results
; no; new
yes
C124
Function group
CURRENT OUTPUT
O
Selection of current output
Out2
Out1;
O1
O2
Selection characteristic
Tab = Table
sim = Simulation
lin = linear
O3 (3)
O3 (2)
O3 (1)
Entry of simulation value
0 ... 22.00 mA
current value
O321
Function group
CHECK
P
Selection of alarm
threshold monitoring
, Low, High,
Lo+Hi = lower and upper
off
P1
P3
Setting of lower
alarm threshold
0 ... 19.00 mg/l
0 ... 190.0 %SAT
0.00 mg/l;
0.0 %SAT;
P4
Setting of upper
alarm threshold
1 ... 20.00 mg/l
10.0 ... 200.0 %SAT
20.00 mg/l;
200.0 %SAT;
Selection of
process monitoring
, AC, CC, AC+CC
off
Monitoring limit
AC:
0.1 mg/l
1 %SAT
±
±
P5
Alarm delay
0 ... 2000 min (s)
(depends on F2)
0 min (s)
P2
Selection of table option
edit
read
O331
Entry of number of
value pairs in table
1 ... 10
1
O332
Selection of value
pair in table
1 ... number of value
pair in table
asign
1
O333
Selection of
current range
0-20 mA
4-20 mA;
O311
O312
Entry of 20 mA value
0 ... 20.00 mg/l
0 ... 200.0 %SAT
10.00 mg/l;
100.0 %SAT;
40.0 °C; –10.0 ... 60.0 °C
O313
Entry of 0/4 mA value
0...20.00 mg/l
0...200.0 %SAT
0.00 mg/l;
0.0 %SAT;
0.0 °C; –10.0 ... 60.0 °C
Function group
ALARM
F
Selection of contact type
=
Fleet = fleeting contact
Stead
Steady cont.
F1
Selection of
alarm delay unit
; s
min
F2
Alarm delay
0 s ... 2000 min (s)
(depends on F2)
0 min (s)
F3
Setting of
error current
2.4 mA
22 mA;
F4
Selection of
error number
1 ... 255
1
F5
Set alarm contact
to be effective
; no
yes
F6
Setting of maximal
permissible limit
violation period
0 ... 2000 min
480 min
P6
Funktionsgruppe
STROMEINGANG
A
Z1
Z2
Z3
Z4
Z5
Z6
Feedforward control
to PID controller
lin = linear
Off;
Cont. switch-off
by current input
Input
Off;
Switch-off direction
for current input
High
Low;
Delay for cont.
switch-off current input
0 ... 2000 s
0 s
Delay for cont.
switch-on current input
0 ... 2000 s
0 s
Switch-off limit value
for current input
0 ... 100%
50%
Selection of meas. unit for
2nd current output
; mg/l; Sat; %; Contr
°C
Appendix
Liquisys M COM 223 F / 253 F
72