Error
no.
Cause
Measures
Contact
Error current
Automatic
cleaning trigger
Fact
User
Fact
User
Fact
User
E081
Distance between
0/4 mA and 20 mA values
at output 2 too short
Adjust spreading for current output 2.
no
no
—
—*
E082
Air pressure range
undershot (< 500 mbar)
Measurement only possible when
uncompensated
or with location height data.
yes
no
no
E083
Air pressure range
exceeded (>1100 mbar)
yes
no
no
E100
Current simulation active
no
no
—
—*
E101
Service function yes
Switch off service function or switch
instrument off and back on.
no
no
—
—*
E102
Manual mode active
no
no
—
—*
E106
Download yes
Wait for download to end.
no
no
—
—*
E116
Download error
Repeat download.
no
no
—
—*
E152
Meas. signal of main
parameter sluggish or
frozen
Check sensor and connecting line, service
or replace.
yes
no
no
E154
Below lower alarm
threshold for period
exceeding alarm delay
Perform manual reference measurement if
necessary. Service sensor and recalibrate.
yes
no
no
E155
Above upper alarm
threshold for period
exceeding alarm delay
yes
no
no
E156
Actual value undershoots
monitoring value for more
than permissible
maximum period
yes
no
no
E157
Actual value exceeds
monitoring value for more
than permissible
maximum period
yes
no
no
E162
Dosage stop
Check settings in CURRENT INPUT or
CHECK function group.
yes
no
no
E171
Flow in main stream loo
low or zero
Restore flow.
yes
no
no
E172
Switch-off limit for current
iput exceeded
Check process variables at sending
measuring instrument.
yes
no
no
E173
Current input < 4 mA
Check process variables at sending
measuring instrument.
Change range assignment if required.
yes
no
no
E174
Current input > 20 mA
Check process variables at sending
measuring instrument.
Change range assignment if required.
yes
no
no
* When this error is present, the cleaning function cannot be started.
(Field F8 does not exist for this error.)
Maintenance and troubleshooting
Liquisys M COM 223 F / 253 F
56