Coding
Field
Selection or range
default (bold)
Display
Info
R211
Switch
function of R2 (1)
off or on
Off
On
All settings are retained.
R212
Enter switch-on
point of contact
20.00 mg/l
0.00 ... 20.00 mg/l
200.0 %SAT
0.0 ... 200.0 %SAT
Never set switch-on point and
switch-off point to the same value!
(The unit selected in A1 appears.)
R213
Enter contact
switch-off point
20.00 mg/l
0.00 ... 20.00 mg/l
200.0 %SAT
0.0 ... 200.0 %SAT
Entering the switch-off point
selects either a max contact
(switch-off point < switch-on
point) or a min contact (switch-off
point > switch-on point) and
implements a constantly required
hysteresis (see Fig. 5.4).
R214
Enter pick-up delay
0 s
0 ... 2000 s
R215
Enter dropout
delay
0 s
0 ... 2000 s
R216
Enter alarm
threshold (as an
absolute value)
20.00 mg/l
0.00 ... 20.00 mg/l
200.0 %SAT
0.0 ... 200.0 %SAT
When the alarm threshold is
violated, the measuring
transmitter issues an alarm with
an error message and error
current (note alarm delay).
When defining the min contact,
the alarm threshold must be set
to a lower value than the
switch-off point.
R217
Show status for
limit contact
MAX
MIN
Only display.
R2 (2)
Configure limit
contactor for
temperature
measurement
LC PV =
limit contactor O
2
(1)
LC °C
=
limit contactor T
(2)
PID controller (3)
Timer (4)
Clean = ChemoClean (5)
Confirmation with ENTER
switches off a different, already
switched-on function and its
settings are reset to the default.
The factory settings are printed in bold; functions printed in
italics are not available in the basic version.
Off
R211
Function
20.00
mg/l
R212
On value
20.00
mg/l
R213
Off value
0
s
R214
On Delay
0
s
R215
Off Delay
20.00
mg/l
R216
A.Thresh
R237
LC State
MIN
LC °C
R2
Sel. Type
OM23FE5B.CHP
Liquisys M COM 223 F / 253 F
Instrument configuration
Hauser