Operation
Liquistation CSF39
84
Hauser
Settings with the Basic program type with 2 bottles
Sampling mode
=
Flow paced VTCV
Menu/Setup/Sampling programs/Setup program/New/Basic
Function
Options
Info
Flowmeter input
Selection
• No flow input
configured
• Binary input S:x
• Current input S:x
Factory setting
No flow input configured
Select the flow input. The binary input or the
current input must be configured for this
function. Only the inputs configured as a
flow input are displayed.
Sampling interval
(for version
with vacuum pump or peristaltic
pump)
Sampling interval
1000 to 9,999,000 m
3
3
Factory setting
10,000 m
3
3
Set the sampling interval. The unit and the
number of decimal places are displayed as
configured under
Setup
/
Inputs
.
Dosing volume
(for version with
vacuum pump or sampling
assembly)
Sampling volume
(for version with
peristaltic pump)
Sampling volume
Factory setting
Vacuum pump:
20 to 350 ml
Peristaltic pump:
10 to 10000 ml
Sampling assembly:
10 to 1000 ml
Factory setting
Vacuum pump:
200 ml
Peristaltic pump:
100 ml
Sampling assembly:
200 ml
Set the dosing volume or the sampling
volume.
Set the sampling volume.
In the version with a vacuum pump or
sampling assembly, the volume is taken
from the setup and can only be modified
there.
The dosing accuracy and the repeatability of
a sample volume < 20 ml can vary,
depending on the specific application.
Multiplier
(for version with vacuum pump or
sampling assembly)
1 to 10
Factory setting
1
You can change the sampling volume using
the multiplier. For example, if the dosing
volume is set to 200 ml, set the multiplier to
2 for a sampling volume of 400 ml. 2
samples are taken in succession when
sampling.
Bottle change mode
Selection
• Number of samples
• Time
• External signal
Factory setting
External signal
The bottle can be changed after a specific
number of samples, after a time or by an
external signal.
If is selected:
Bottle change mode
Number of samples
Samples per bottle
1 ... 9999
Factory setting
1
Set the number of samples.
Having selected
Bottle change mode
Time
Time interval
00-00:02 ... 31-00:00
DD-HH:MM
Factory setting
00
‐
01:00 DD
‐
HH:MM
Set the time (days, hours and minutes) after
which the system should change to the next
bottle.
Multiple bottles
0 ... 23
The configuration options
depend on the current
number of bottles
Factory setting
0
Multiple bottles
"Simultaneous" transfer of two samples to
separate bottles.