Operation
Liquistation CSF39
60
Hauser
Menu/Setup/General settings/Logbooks
Function
Options
Info
Source of data
Read only
This is for information purposes only. If you
want to record another value, delete this logbook
and create a new data logbook.
Measured value
Log time left
Event logbook
=
Fill up buffer
Read only
Displays the days, hours and minutes remaining
until the logbook is full.
Log size
Event logbook
=
Fill up buffer
Read only
Displays the number of entries remaining until
the logbook is full.
Logbook name
Customized text, 20
characters
You can change the name here again.
Scan time
0:00:01 to 1:00:00
Factory setting
0:01:00
As above
Minimum time interval between two entries
Format: H:MM:SS
Data logbook
Selection
• Ring buffer
• Fill up buffer
Factory setting
Ring buffer
Ring buffer
If the memory is full, the most recent entry
automatically overwrites the oldest entry.
Fill up buffer
If the memory is full, there is an overflow, i. e. no
new values can be saved. The controller displays
a corresponding diagnostic message. The
memory then has to be cleared manually.
Overflow warnings
Event logbook
=
Fill up buffer
Selection
• Off
• On
Factory setting
Off
‣
Decide whether you want to receive a
diagnostic message if the fill buffer of the
relevant logbook overflows.
Line plotter
Menu to define the graphic display
Axes
Selection
• Off
• On
Factory setting
On
Should the axes (x, y) be displayed (
On
) or not
(
Off
)?
Orientation
Selection
• Horizontal
• Vertical
Factory setting
Horizontal
You can choose whether the value curves should
be displayed from left to right (
Horizontal
) or
from top to bottom (
Vertical
). If you want to
display two data logbooks simultaneously, make
sure that both logbooks have the same settings
here.
X-Description
Selection
• Off
• On
Factory setting
On
‣
Decide whether a description should be
displayed for the axes and whether gridlines
should be shown. In addition, you can also
decide whether pitches should be displayed.
Y-Description
Grids
Pitches
X Pitch/Grid distance
10 to 50%
Factory setting
10 %
‣
Determine the pitch.
Y Pitch/Grid distance
Remove
Action
This action removes the data logbook. Any data
that have not been saved are lost.
Example: New data logbook (Setup/General settings/Logbooks/Data logbooks/New)
1. Make the settings: