Operation
Liquistation CSF39
58
Hauser
You define how the logbooks should store the data.
In addition, you can also define individual data logbooks .
1. Assign the logbook name.
2. Select the measured value to be recorded.
3. Set the scan time (
Scan time
).
You can set the scan time individually for every data logbook.
Further information on the logbooks: .
Menu/Setup/General settings/Logbooks
Function
Options
Info
Logbook ident
Customized text, 16
characters
Part of the file name when exporting a logbook
Event logbook
Selection
• Off
• Ring buffer
• Fill up buffer
Factory setting
Ring buffer
All diagnostic messages are recorded
Ring buffer
If the memory is full, the most recent entry
automatically overwrites the oldest entry.
Fill up buffer
If the memory is full, there is an overflow,i. e.
you cannot store any new values. The controller
displays a corresponding diagnostic message.
The memory then has to be cleared manually.
Logbook program
Selection
• Off
• Ring buffer
• Fill up buffer
Factory setting
Ring buffer
All program cycles are recorded
Ring buffer
If the memory is full, the most recent entry
automatically overwrites the oldest entry.
Fill up buffer
If the memory is 80 % full, the device displays a
diagnostic message.
If the memory is full, there is an overflow, i. e. no
new values can be saved. The controller displays
a corresponding diagnostic message. The
memory then has to be cleared manually.
Overflow warnings
Event logbook
=
Fill up buffer
Calibration logbook
Selection
• Off
• On
Factory setting
Off
‣
Decide whether you want to receive a
diagnostic message if the fill buffer of the
relevant logbook overflows.
Diagnostic logbook
Configuration logbook