51
pH
Analog pH/mV input
pH socket, DIN 19 262 (13/4 mm)
pH range
-2 … 16
Decimal places *
)
2 or 3
Input resistance
1 x 10
12
Ω
(0 … 35 °C)
Input current
1 x 10
-12
A
(at RT, doubles every 10 K)
Measuring cycle
Approx. 1 s
Measurement error
1,2,3)
< 0.01 pH, TC < 0.001 pH/K
mV range
-1300 … +1300 mV
Measuring cycle
Approx. 1 s
Measurement error
1,2,3)
< 0.1 % meas. val. + 0.3 mV, TC < 0.03 mV/K
* User-defined
1) according to EN 60746-1, at nominal operating conditions
2) ± 1 count
3) plus sensor error
4) Ranges depending on Memosens sensor
Memosens pH input
(also ISFET)
M8 socket, 4 pins, for Memosens lab cable or
M12 socket for Memosens sensors
Display ranges
4)
pH
-2.00 … +16.00
mV
-1999 … +1999 mV
Temperature
-50 … +250 °C (-58 ... +482 °F)
Memosens input
ORP
M8 socket, 4 pins, for Memosens lab cable or
M12 socket for Memosens sensors
Display ranges
4)
mV
-1999 … +1999 mV
Temperature
-50 … +250 °C (-58 ... +482 °F)
Sensor standardization *
ORP calibration (zero adjustment)
Permissible calibration range
ΔmV (offset)
-700 … +700 mV
Specifications