52
pH
Specifications
Sensor standardization *
pH calibration
Operating modes *
Programmed buffers
Calibration with automatic buffer recognition
Manual
Manual calibration with entry of individual
buffer values
Data entry
Data entry of zero and slope
Buffer sets *
-01- Hauser
2.00/4.01/6.98/9.95/11.87
-02- Mettler-Toledo
2.00/4.01/7.00/9.21
-03- Knick CaliMat
2.00/4.00/7.00/9.00/12.00
-04- Ciba (94)
2.06/4.00/7.00/10.00
-05- NIST technical
1.68/4.00/7.00/10.01/12.46
-06- NIST standard
1.679/4.006/6.865/9.180
-07- HACH
4.01/7.00/10.01/12.00
-08- WTW techn. buffers 2.00/4.01/7.00/10.00
-09- Hamilton
2.00/4.01/7.00/10.01/12.00
-10- Reagecon
2.00/4.00/7.00/9.00/12.00
-11- DIN 19267
1.09/4.65/6.79/9.23/12.75
Permissible calibration range
Zero point
6 ... 8 pH
With ISFET: Operating
point (asymmetry)
-750 … +750 mV
Slope
Approx. 74 … 104 %
(possibly restricting notes from Sensoface)
Calibration timer
*
Interval 1 … 99 days, can be switched off
Sensoface
Provides information on the sensor condition
Evaluation of
zero/slope, response, calibration interval
* User-defined