29
Cond
”Entry of Solution” Calibration
(Calibration by entering the conductivity with display of the cell constant)
1) Press the Start softkey.
2) Immerse the sensor in the solution.
3) Enter the temperature-corrected conductivity value and press E.
4) Finally, you can Apply these values or Cancel the calibration.
Cell Constant Calibration
(Calibration by entering the cell constant with display of the conductivity)
1) Press the Start softkey.
2) Immerse the sensor in the solution.
3) Modify the value of the cell constant until the temperature-corrected
conductivity value is reached. Then press E.
4) Finally, you can Apply these values or Cancel the calibration.
Conductivity sensor
Cell constant
CLS16D
k=0.1 cm
-1
CLS21D
k=1.0 cm
-1
Calibration
Note: To abort calibration, you can press
at any time.