21
Cond
Configuration
Conductivity Configuration
1) In measuring mode, press the Menu softkey.
2) Select “Configuration” and confirm by pressing E.
3) Make the desired adjustments.
The following table gives you an overview. Factory settings are shown in bold print.
Language
E
Deutsch | English | Español | Italiano | Français |
Português
Auto-off
Off | 5 min | 10 min | 30 min | 60 min
Temperature
°C | °F
+ Cond sensor*
Conductivity
S/cm | S/m
Range selection
Auto | o.ooo µS/cm | oo.oo µS/cm |
ooo.o µS/cm | oooo µS/cm |
oo.oo mS/cm | ooo.o mS/cm | oooo mS/cm
Calculation
Off | MΩ cm | TC | SAL | TDS
TC compensation
TC: Linear | NLF | NaCl | HCl | NH3 | NaOH
TC of solution
TC: 0 … 20.0 %/K | 1.00 %/K
Ref. temp
TC: 0 … 100.0 °C | 25 °C 32 … 212 °F | 77 °F
TDS factor
TDS: 0 …9.99 | 1.00
“Configuration” menu selection – part 1