26
pH
Calibration
(Automatic calibration with specification of the buffer solution used)
1) Select the number of calibration points and the buffer set as shown in the table
below and press the Start softkey.
Calibration points Auto | 1-point | 2-point | 3-point
Hauser
2.00/4.01/6.98/9.95/11.87
Buffer set
Mettler-Toledo
2.00/4.01/7.00/9.21
Knick CaliMat
2.00/4.00/7.00/9.00/12.00
Ciba
2.06/4.00/7.00/10.00
NIST technical
1.68/4.00/7.00/10.01/12.46
NIST standard
1.679/4.006/6.865/9.180
Hach
4.01/7.00/10.01/12.00
WTW
2.00/4.01/7.00/10.00
Hamilton
2.00/4.01/7.00/10.01/12.00
Reagecon
2.00/4.00/7.00/9.00/12.00
DIN 19267
1.09/4.65/6.79/9.23/12.75
2) Immerse the sensor in the 1st/2nd/3rd buffer solution and press Continue
(repeat this step for each calibration point).
3) Finally, the calibration data will be displayed. You can Apply or Discard these
values.
pH Calibration
1) In measuring mode, press the Menu softkey.
2) Select “Configuration” and confirm by pressing E.
3) Select the desired “Calibration mode”.
4) Perform the selected calibration as described on the following pages.
Follow the instructions on the display.
Calibration
Note: To abort calibration, you can press
at any time.