50
pH
Oxy Cond
Specifications
Communication
USB 2.0
Profile
HID, driverless installation
Usage
Data exchange
Diagnostics functions
Sensor data
(Memosens only)
Manufacturer, sensor type, serial number, wear, operating time
Calibration data
Calibration date; pH/Oxy: zero, slope; Cond: Cell constant
Device self-test
Automatic memory test (FLASH, EEPROM, RAM)
Device data
Device type, software version, hardware version
Data retention
Parameters, calibration data > 10 years
EMC
EN 61326-1 (General Requirements)
Emitted interference
Class B (residential area)
Immunity to interference
Industry
EN 61326-2-3 (Particular Requirements for Transmitters)
RoHS conformity
According to directive 2011/65/EU
Power supply
4 x AA alkaline batteries or
4x rechargeable NiMH batteries
1x Li-ion battery, USB chargeable
Nominal operating conditions
Ambient temperature
-10 … +55 °C (+14 ... +130 °F)
Transport/
Storage temperature
-25 … +70 °C (-13 ... +158 °F)
Relative humidity
0 … 95 %, short-term condensing allowed
Housing
Material
PA12 GF30 (silver gray RAL 7001) + TPE (black)
Protection
IP66/67 with pressure compensation
Dimensions
Approx. (132 x 156 x 30) mm
Weight
Approx. 500 g