19
pH
pH Configuration
1) In measuring mode, press the Menu softkey.
2) Select “Configuration” and confirm by pressing E.
3) Make the desired adjustments.
The following table gives you an overview.
Factory settings are shown in bold print.
“Configuration” menu selection – part 1
Language
E
Deutsch | English | Español | Italiano | Français |
Português
Auto-off
Off | 5 min | 10 min | 30 min | 60 min
Temperature
°C | °F
+ pH sensor*
Wear
On | Off
+ Calibration*
Cal timer
Off | On
Interval
On: 00 … 99 days
Cal mode
Programmed buffer | Manual | Data entry
Calibration
points
Auto | 1-point | 2-point | 3-point
Buffer set
Hauser
Mettler-Toledo
Knick CaliMat
Ciba
NIST technical
NIST standard
Hach
WTW
Hamilton
Reagecon
DIN 19267
2.00/4.01/6.98/9.95/11.87
2.00/4.01/7.00/9.21
2.00/4.00/7.00/9.00/12.00
2.06/4.00/7.00/10.00
1.68/4.00/7.00/10.01/12.46
1.679/4.006/6.865/9.180
4.01/7.00/10.01/12.00
2.00/4.01/7.00/10.00
2.00/4.01/7.00/10.01/12.00
2.00/4.00/7.00/9.00/12.00
1.09/4.65/6.79/9.23/12.75
MemoLog
Off | On
TAG
Off | On
Configuration
* “+” indicates that submenus can be opened by pressing E.