Operation
Liquiline System CA80AL
82
Hauser
Menu/Setup/Outputs/Binary output x:y
Function
Options
Info
Binary output
Selection
On
Signal type
Selection
Static signal
Function
Selection
Analyzer
Assignments
Function
=
Analyzer
Selection
Measurement active SP1
Here select which binary outputs output the
system status for a measurement currently
running.
In the case of two-channel devices,
Measurement active SP2
is also displayed
1)
x:y = slot no. : input number
9. Press
OK
to confirm.
The binary inputs and binary outputs are configured.
10. Switch to the Automatic mode: press the soft key
MODE
and select
Continue
automatic mode
or
Start automatic mode
.
The display shows
Current mode
-
Automatic
.
Activating an external pump before every measurement
If you convey the sample to the analyzer using an external pump or an external sample
preparation system, you can use the
Signal for sample request
to switch on the external
devices for a limited time only when the analyzer requires sample. The signal becomes
active before every measurement, calibration and cleaning. You can configure how long
the signal should be active for. The start of the actual analyzer activity is delayed by this
time.
1. Under
Menu
/
Setup
/
Analyzer
/
Extended setup
/
Signal for sample request
must be
selected.
2. Under
Lead time
set how long an action that requires sample (measurement,
calibration or cleaning) is delayed.
3. Under
Duration SP%C
set how long the signal should remain active. The duration
can be longer than the lead time. The maximum possible value equals the lead time
plus the duration of the measurement.
4. Configure the binary outputs as follows:
Menu/Setup/Outputs/Binary output x:y
Function
Options
Info
Binary output
Selection
On
Signal type
Selection
Static signal
Function
Selection
Analyzer
Assignments
Function
=
Analyzer
Selection
Sample required SP1
Here select which binary outputs output the
system status for a measurement currently
running.
In the case of two-channel devices,
Sample
required SP2
is also displayed
1)
x:y = slot no. : input number