Liquiline System CA80AL
Technical data
Hauser
163
15
Technical data
15.1 Input
Measured variables
Al [µg/l, ppb]
Measuring range
15 to 1000 µg/l (ppb) Al
Types of input
• 1 or 2 measuring channels (analyzer main parameter)
• 1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)
• Analog current inputs (optional)
• Binary inputs (optional)
Input signal
Depending on version
2 x 0/4 to 20 mA (optional), passive, potentially isolated
Current input, passive
Span
> 0 to 20 mA
Signal characteristic
Linear
Internal resistance
Non-linear
Test voltage
500 V
Hose specification (self-
priming analyzer)
• Clearance: max. 1.0 m (3.3 ft)
• Height: max. 0.5 m (1.6 ft)
• Hose ID: 1.6 mm (¹/
16
in)
Cable specification
(for optional sensors with
Memosens technology)
Cable type
Memosens data cable CYK10 or sensor fixed cable, each with cable end sleeves or M12
round-pin connector (optional)
Cable length
Max. 100 m (330 ft)
15.2 Output
Output signal
Depending on version:
• 2 x 0/4 to 20 mA, active, potentially isolated (standard version)
• 4 x 0/4 to 20 mA, active, potentially isolated (version with 2 additional analog outputs)
• 6 x 0/4 to 20 mA, active, potentially isolated (version with 4 additional analog outputs)
• Binary outputs
Modbus RS485
Signal encoding
EIA/TIA-485
Data transmission rate
2,400, 4,800, 9,600, 19,200, 38,400, 57,600 and 115,200
baud