Maintenance
Liquiline System CA80AL
146
Hauser
Using the standard calibration solution again
1. Replace the zero standard with the standard. Use either an additional hose in order
to reconnect the standard calibration solution, or allow the hose to drain and dry the
hose well.
2. Connect the standard calibration solution to the Liquid Manager.
3. Select
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Bottle selection
.
4. Select
Standard S1
and select
OK
to confirm.
5. Select
Confirm bottles inserted
.
Recommendation: Afterwards, perform a manual calibration with the standard
solution (
Determine calibration factor
).
Alternatively, the zero point and calibration factor can be determined in one step
(
Determine zero point and calibr. factor
).
12.5 Replacing the hoses
You require the following parts:
• From the CAV800 maintenance kit:
• NORPRENE hose, ID 1.6 mm
• C
‐
Flex hose, ID 3.2 mm
• C
‐
Flex hose, ID 6.4 mm
• Hose connector
• 1 pair of gloves that are resistant to the reagents used
1. Press the
MODE
soft key and select the
Manual mode
.
The display shows
Current mode
-
Manual
.
Wait until all operations have stopped. Ongoing operations can be stopped under
Menu
/
Operation
/
Manual operation
.
There should not be any more sample left in the sample collecting vessel.
3. Remove the bottles and the bottle tray.
4. Remove the cover of the carrier board.
5. Replace with hoses of the same diameter and length. Attach hose labels to the new
hoses.
6. Route outlet hose D6 behind dispenser 7.
7. Secure the cover and place the bottle tray back into the housing.
8. Connect the bottles to the appropriate covers and hoses.
9. Select
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Bottle selection
.
10. Select all the bottles and press
OK
to confirm.
11. Select
Confirm bottles inserted
.
12. Press the
MODE
soft key and select
Continue automatic mode
or
Start automatic
mode
.
12.6 Replacing filter mats
You require the following parts:
Filter mats (part of the CAV800 maintenance kit)
Summary of Contents for Liquiline System CA80AL